Clamp-on sensors
The sensors strap to the outside of the pipe, so the pipe stays closed: no cutting and no draining. The pipe must be full of liquid where the sensors sit.
Technical details method, spacing, mounting
- Method
- Transit time on a V path: both sensors sit on one side of the pipe, and the sound reflects once off the far wall.
- Sensor spacing
- Set from the outside diameter, wall thickness, pipe material, lining and liquid.
- Contact
- Clean the spot, remove loose paint or rust, and use coupling gel between sensor and pipe.
- Position
- A straight run away from bends, valves and pumps. On horizontal pipes, mount at the side: air collects at the top and sediment at the bottom.
- Lined pipe
- Support per model: not confirmed yet. Ask an engineer.
- Insertion
- Also available as insertion versions, which go through the pipe wall: D118, E5E.
General guidance for clamp-on transit-time meters. Distances and limits depend on the model.